Monocyte Activation Test (MAT) as a possibility of replacement for the rabbit pyrogen test in hyperimmune sera
Autor
Afiliação Butantan
Afiliação externa
Tipo de documento
Article
Idioma
English
Direitos de acesso
Open access
Licença de uso
CC BY
Aparece nas Coleções:
Métricas
Resumo em inglês
The use of a commercial kit for the monocyte-activation test (MAT) was evaluated for assessing pyrogenic contamination of hyperimmune sera. Three batches of sera, two pyrogen free and one pyrogenic, were tested. Endotoxin spike recover indicated that sample dilutions from 1/2 to 1/10 are suitable. Kit transport and storage conditions were also evaluated, proving that an adequate cold chain must be assured to achieve good results. Furthermore, the commercial MAT kit seemed suitable to replace the rabbit pyrogen test (RPT) for pyrogen testing of hyperimmune sera, although further tests are needed to a full validation.
Referência
Utescher CLA, Buosi KL, Botosso VF, Quintilio W. Monocyte Activation Test (MAT) as a possibility of replacement for the rabbit pyrogen test in hyperimmune sera. Braz. J. Pharm. Sci.. 2018;54(2):e17530. doi:10.1590/s2175-97902018000217530.
URL permanente para citação desta referência
https://repositorio.butantan.gov.br/handle/butantan/2543
URL
http://dx.doi.org/10.1590/s2175-97902018000217530
Sobre o periódico
Palavra-chave
Agência de fomento
Data de publicação
2018
Arquivos neste item
Este item está licenciada sob uma Licença Creative Commons